Abstract
A 74-year-old pacemaker-dependent male patient, who underwent a pacemaker generator change due to elective replacement indicator, had a 4-second pause during interrogation of the new pacemaker generator out of the device pocket due to a specific feature that was programmed on called "enhanced transtelephonic monitoring." (PACE 2009; 32:949-951)
Original language | English (US) |
---|---|
Pages (from-to) | 949-951 |
Number of pages | 3 |
Journal | PACE - Pacing and Clinical Electrophysiology |
Volume | 32 |
Issue number | 7 |
DOIs | |
State | Published - Jul 2009 |
Keywords
- Pacing
- Transtelephonic monitoring