Influence of hip prostheses on high energy photon dose distributions

C. H. Sibata, H. C. Mota, P. D. Higgins, D. Gaisser, J. P. Saxton, K. H. Shin

Research output: Contribution to journalArticlepeer-review

39 Scopus citations

Abstract

Radiotherapy treatment of patients having a hip prosthesis is a common problem facing dosimetrists and physicists when the treatment plan requires irradiation of the pelvic area. To quantify the perturbation of these devices, attenuation studies were done with 6 and I8 MV photon beams using various hip prostheses models with varying size and composition. These studies have shown that an attenuation of as much as 50% can be found in a single beam profile under the prosthesis. We have studied the capability of a dose planning system to predict the transmission of these devices as compared with measurements.