Abstract

Objective: There is increased confusion regarding MRI-compatible CIs and BAHAs. This report describes two cases when patients underwent MRIs with non-MRI compatible devices. Results: One patient with bilateral Cochlear Osias experienced dislocation of both internal magnets after 1.5 Tesla MRI. Both magnets were outside the silastic sheath, with the left magnet flipped. A second patient with a legacy CI experienced similar internal magnet dislocation and inversion after 3 Tesla MRI. Conclusions: This study describes internal magnet dislocation/inversion with the Cochlear Osia and a legacy CI after MRI. Our findings suggest the need for improved patient education and simplified radiology guidelines. Laryngoscope, 134:393–396, 2024.

Original languageEnglish (US)
Pages (from-to)393-396
Number of pages4
JournalLaryngoscope
Volume134
Issue number1
DOIs
StatePublished - Jan 2024

Bibliographical note

Publisher Copyright:
© 2023 The Authors. The Laryngoscope published by Wiley Periodicals LLC on behalf of The American Laryngological, Rhinological and Otological Society, Inc.

Keywords

  • MRI safety
  • adverse events
  • cochlear implant
  • osseointegrated implant

PubMed: MeSH publication types

  • Case Reports

Fingerprint

Dive into the research topics of 'Magnetic Resonancy Imaging Safety in Active Osseointegrated (Osia®) and Cochlear Implants: New Technology Creating Confusion'. Together they form a unique fingerprint.

Cite this