TY - JOUR
T1 - Measurement of intraocular pressure after epikeratophakia
AU - Wright, Martha M.
AU - Grajewski, Alana L.
PY - 1997/6
Y1 - 1997/6
N2 - Aims - To assess the accuracy of three commonly used tonometers in eyes after epikeratophakia. Methods - Five eye bank eyes with sutured epikeratophakia buttons were connected to a manometer and a pressure transducer. Intraocular pressure was adjusted in 5 mmHg increments from 0 to 50 mmHg. The intraocular pressure was measured at each increment using a Goldmann tonometer, a pneumatonometer, and a Tono-pen. Results - The difference between the manometer (actual pressure) and the Goldmann tonometer ranged from -19 to +9 mmHg (mean (SD) overestimation 2.6 (5.8) mmHg). The pneumatonometer error ranged from -27.5 to +5.5 mmHg (mean (SD) overestimation 4.7 (6.1) mmHg), and for the Tono-pen the range was -18 to +11 mmHg (mean (SD) overestimation 0.05 (7.9) mmHg). The correlation coefficients for the three tonometers were 0.94, 0.92, and 0.87 for the Goldmann tonometer, pneumatonometer, and Tono-pen respectively. Conclusion - The Goldmann tonometer had the best correlation with the manometer in eye bank eyes with epikeratophakia (correlation coefficient 0.94), but none of the tonometers was accurate over the entire range of pressures tested. Detection of glaucoma in eyes with epikeratophakia cannot rely on tonometry alone, but requires examination of the optic nerve and visual field.
AB - Aims - To assess the accuracy of three commonly used tonometers in eyes after epikeratophakia. Methods - Five eye bank eyes with sutured epikeratophakia buttons were connected to a manometer and a pressure transducer. Intraocular pressure was adjusted in 5 mmHg increments from 0 to 50 mmHg. The intraocular pressure was measured at each increment using a Goldmann tonometer, a pneumatonometer, and a Tono-pen. Results - The difference between the manometer (actual pressure) and the Goldmann tonometer ranged from -19 to +9 mmHg (mean (SD) overestimation 2.6 (5.8) mmHg). The pneumatonometer error ranged from -27.5 to +5.5 mmHg (mean (SD) overestimation 4.7 (6.1) mmHg), and for the Tono-pen the range was -18 to +11 mmHg (mean (SD) overestimation 0.05 (7.9) mmHg). The correlation coefficients for the three tonometers were 0.94, 0.92, and 0.87 for the Goldmann tonometer, pneumatonometer, and Tono-pen respectively. Conclusion - The Goldmann tonometer had the best correlation with the manometer in eye bank eyes with epikeratophakia (correlation coefficient 0.94), but none of the tonometers was accurate over the entire range of pressures tested. Detection of glaucoma in eyes with epikeratophakia cannot rely on tonometry alone, but requires examination of the optic nerve and visual field.
UR - http://www.scopus.com/inward/record.url?scp=0030746586&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0030746586&partnerID=8YFLogxK
U2 - 10.1136/bjo.81.6.448
DO - 10.1136/bjo.81.6.448
M3 - Article
C2 - 9274407
AN - SCOPUS:0030746586
SN - 0007-1161
VL - 81
SP - 448
EP - 451
JO - British Journal of Ophthalmology
JF - British Journal of Ophthalmology
IS - 6
ER -