The possible influence of third-order shim coils on gradient–magnet interactions: an inter-field and inter-site study

Nicolas Boulant, Caroline Le Ster, Alexis Amadon, Guy Aubert, Alexander Beckett, Jean Belorgey, Cédric Bonnelye, Dario Bosch, David Otto Brunner, Guillaume Dilasser, Olivier Dubois, Philipp Ehses, David Feinberg, Sajjad Feizollah, Vincent Gras, Simon Gross, Quentin Guihard, Hervé Lannou, Denis Le Bihan, Franck MauconduitFrédéric Molinié, François Nunio, Klaas Pruessmann, Lionel Quettier, Klaus Scheffler, Tony Stöcker, Christine Tardif, Kamil Ugurbil, Alexandre Vignaud, An Vu, Xiaoping Wu

Research output: Contribution to journalArticlepeer-review

Abstract

Objective: To assess the possible influence of third-order shim coils on the behavior of the gradient field and in gradient–magnet interactions at 7 T and above. Materials and methods: Gradient impulse response function measurements were performed at 5 sites spanning field strengths from 7 to 11.7 T, all of them sharing the same exact whole-body gradient coil design. Mechanical fixation and boundary conditions of the gradient coil were altered in several ways at one site to study the impact of mechanical coupling with the magnet on the field perturbations. Vibrations, power deposition in the He bath, and field dynamics were characterized at 11.7 T with the third-order shim coils connected and disconnected inside the Faraday cage. Results: For the same whole-body gradient coil design, all measurements differed greatly based on the third-order shim coil configuration (connected or not). Vibrations and gradient transfer function peaks could be affected by a factor of 2 or more, depending on the resonances. Disconnecting the third-order shim coils at 11.7 T also suppressed almost completely power deposition peaks at some frequencies. Discussion: Third-order shim coil configurations can have major impact in gradient–magnet interactions with consequences on potential hardware damage, magnet heating, and image quality going beyond EPI acquisitions.

Original languageEnglish (US)
Pages (from-to)169-183
Number of pages15
JournalMagnetic Resonance Materials in Physics, Biology and Medicine
Volume37
Issue number2
DOIs
StatePublished - Apr 2024

Bibliographical note

Publisher Copyright:
© The Author(s) 2024.

Keywords

  • Field monitoring
  • Gradient
  • Magnet interactions
  • Shim coils
  • Ultra-high field
  • Vibrations

Fingerprint

Dive into the research topics of 'The possible influence of third-order shim coils on gradient–magnet interactions: an inter-field and inter-site study'. Together they form a unique fingerprint.

Cite this