The provision of a uniform vaginal surface dose rate by a novel afterloading cylinder

Jane M Johnson, Roger A. Potish

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

The administration of a uniform dose rate to the vaginal surface is important in the management of endometrial, cervical, and vaginal malignancies. Unfortunately, conventional vaginal cylinders fail to provide this uniformity, and although dome cylinders do so, they require specialized 137Cs sources. Thus, a new acrylic vaginal cylinder has been developed to use with standard 137Cs sources and provides a uniform dose rate independent of vaginal size. Each contoured cylinder follows a particular isodose line. A metal ring is used to secure the device to minimize vulvar trauma associated with other vaginal cylinders. The construction and use of a set of these cylinders has allowed determination of their utility and limitations, which will be discussed in detail. Although these applicators do not completely replace conventional cylinders, they do offer a useful addition to the brachytherapeutic armamentarium.

Original languageEnglish (US)
Pages (from-to)193-198
Number of pages6
JournalMedical Dosimetry
Volume16
Issue number4
DOIs
StatePublished - 1991

Keywords

  • Brachytherapy
  • Gynecologic cancer
  • Intracavitary cesium
  • Vaginal cylinder

Fingerprint

Dive into the research topics of 'The provision of a uniform vaginal surface dose rate by a novel afterloading cylinder'. Together they form a unique fingerprint.

Cite this