TY - JOUR
T1 - Validation of a Zio XT Patch Accelerometer for the Objective Assessment of Physical Activity in the Atherosclerosis Risk in Communities (ARIC) Study
AU - Davoudi, Anis
AU - Urbanek, Jacek K.
AU - Etzkorn, Lacey
AU - Parikh, Romil
AU - Soliman, Elsayed Z.
AU - Wanigatunga, Amal A.
AU - Gabriel, Kelley Pettee
AU - Coresh, Josef
AU - Schrack, Jennifer A.
AU - Chen, Lin Yee
N1 - Publisher Copyright:
© 2024 by the authors.
PY - 2024/2
Y1 - 2024/2
N2 - Background: Combination devices to monitor heart rate/rhythms and physical activity are becoming increasingly popular in research and clinical settings. The Zio XT Patch (iRhythm Technologies, San Francisco, CA, USA) is US Food and Drug Administration (FDA)-approved for monitoring heart rhythms, but the validity of its accelerometer for assessing physical activity is unknown. Objective: To validate the accelerometer in the Zio XT Patch for measuring physical activity against the widely-used ActiGraph GT3X. Methods: The Zio XT and ActiGraph wGT3X-BT (Actigraph, Pensacola, FL, USA) were worn simultaneously in two separately-funded ancillary studies to Visit 6 of the Atherosclerosis Risk in Communities (ARIC) Study (2016–2017). Zio XT was worn on the chest and ActiGraph was worn on the hip. Raw accelerometer data were summarized using mean absolute deviation (MAD) for six different epoch lengths (1-min, 5-min, 10-min, 30-min, 1-h, and 2-h). Participants who had ≥3 days of at least 10 h of valid data between 7 a.m–11 p.m were included. Agreement of epoch-level MAD between the two devices was evaluated using correlation and mean squared error (MSE). Results: Among 257 participants (average age: 78.5 ± 4.7 years; 59.1% female), there were strong correlations between MAD values from Zio XT and ActiGraph (average r: 1-min: 0.66, 5-min: 0.90, 10-min: 0.93, 30-min: 0.93, 1-h: 0.89, 2-h: 0.82), with relatively low error values (Average MSE × 106: 1-min: 349.37 g, 5-min: 86.25 g, 10-min: 56.80 g, 30-min: 45.46 g, 1-h: 52.56 g, 2-h: 54.58 g). Conclusions: These findings suggest that Zio XT accelerometry is valid for measuring duration, frequency, and intensity of physical activity within time epochs of 5-min to 2-h.
AB - Background: Combination devices to monitor heart rate/rhythms and physical activity are becoming increasingly popular in research and clinical settings. The Zio XT Patch (iRhythm Technologies, San Francisco, CA, USA) is US Food and Drug Administration (FDA)-approved for monitoring heart rhythms, but the validity of its accelerometer for assessing physical activity is unknown. Objective: To validate the accelerometer in the Zio XT Patch for measuring physical activity against the widely-used ActiGraph GT3X. Methods: The Zio XT and ActiGraph wGT3X-BT (Actigraph, Pensacola, FL, USA) were worn simultaneously in two separately-funded ancillary studies to Visit 6 of the Atherosclerosis Risk in Communities (ARIC) Study (2016–2017). Zio XT was worn on the chest and ActiGraph was worn on the hip. Raw accelerometer data were summarized using mean absolute deviation (MAD) for six different epoch lengths (1-min, 5-min, 10-min, 30-min, 1-h, and 2-h). Participants who had ≥3 days of at least 10 h of valid data between 7 a.m–11 p.m were included. Agreement of epoch-level MAD between the two devices was evaluated using correlation and mean squared error (MSE). Results: Among 257 participants (average age: 78.5 ± 4.7 years; 59.1% female), there were strong correlations between MAD values from Zio XT and ActiGraph (average r: 1-min: 0.66, 5-min: 0.90, 10-min: 0.93, 30-min: 0.93, 1-h: 0.89, 2-h: 0.82), with relatively low error values (Average MSE × 106: 1-min: 349.37 g, 5-min: 86.25 g, 10-min: 56.80 g, 30-min: 45.46 g, 1-h: 52.56 g, 2-h: 54.58 g). Conclusions: These findings suggest that Zio XT accelerometry is valid for measuring duration, frequency, and intensity of physical activity within time epochs of 5-min to 2-h.
KW - ActiGraph GT3X
KW - validation
KW - wearables
KW - Zio XT Patch
UR - http://www.scopus.com/inward/record.url?scp=85184663245&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85184663245&partnerID=8YFLogxK
U2 - 10.3390/s24030761
DO - 10.3390/s24030761
M3 - Article
C2 - 38339479
AN - SCOPUS:85184663245
SN - 1424-8220
VL - 24
JO - Sensors
JF - Sensors
IS - 3
M1 - 761
ER -